Before the Analysis¶
Sample Preparation¶
Filter samples before injection with a
Syringe
andFilter
.- Syringe filters can be found in the cupboard in the weighing room.
- Clean syringes can be found in sealed packets in the cupboard under the
HPLC
.
Pre-Run¶
Check the indicator lights on the
Autosampler
,Pump
,Column
,VWD
andMS
are notRed
.Orange
,Flashing Orange
orOff
are OK.Check the valve near the
ESI
on theMS
is open (handle in line with pipe) and is connected toL
on the nearby block (shown below)Check the levels of the bottles on the front of the
MS
. Tap the bottles to disturb the liquid if the level is not obvious.Check the levels of the waste bottles in the following locations:
- Beneath the bench below the
Pump
/Autosampler
/Eluent
stack - Beneath the bench below the
MS
and behind theVacuum Pumps
- Behind the
MS
- Beneath the bench below the
Check there is pressure in the
Collision Gas Bottle
behind theMS
, and record the pressureCheck the 3
Exhaust Pipes
are entering theExtractor
for theGraphite Furnace
Return to the front of the instrument
Fetch some blue towel.
Open the
Ionisation Chamber
of theMS
and spray50% IPA
around (but not into) the metal surfaces of the inlet orifice.Spray the inside of the
Ionisation Chamber
door and wipe with the blue towel.Close the
Ionisation Chamber
.Open Mass Hunter Workstation Data Acquisition.
Check the pressures on the top right of the screen are within the following values:
Rough Vac < 4.4×10 0 Quad Vac 1×10 −5 - 4×10−5 TOC Vac 2×10 −7 - 4×10 −7 Check sufficient eluent is present in the eluent bottles being used for the run.
Check that the
Quat. Pump
is aware of the level of eluent in the bottles. To do this:Right click on the
Quat. Pump
panel of the Instrument Status window in MassHunter. <could use image here>Select Bottle Fillings from the menu.
Under Actual Volume, ensure the value reflects the volume of eluent actually present in the bottle.
Click OK.
Repeat the above steps when Changing Eluent (see below)
Changing Eluent¶
To change an eluent, perform the following steps:
Right click on the
QTOF
panel of the Instrument Status window in MassHunter and select LC → Waste from the menu.In the Method Editor, under
Quat. Pump
, set the Flow Rate to 0.000 mL/min and press Enter⏎Swap the the eluent bottles. Try to swap similar eluents; for example, replace Ammonium Formate with Ammonium Acetate.
- The Lids on the bottles do not need to be tight.
Open the valve behind the door on the front of the
Pump
.Set the level of any solvents that weren’t changed to 0% and enable the eluents that were changed.
Set the Flow Rate to 1.000 mL/min and press Enter⏎
Check the pressure in the
Column
does not rise above 0 Bar. If it does, check the valve at the front of thePump
is open.Set the Flow Rate to 5.000 mL/min and press Enter⏎
After 5 minutes, set the Flow Rate to 0.000 mL/min and press Enter⏎
Close the valve on the front of the
Pump
and close the door.
A method can be setup for flushing and equilibrating the column as part of the Worklist if desired