Mass Calibration / Check

  1. Select ContextTune at the top left of the MassHunter window.

  2. Turn on the QTOF by clicking the On button in its section of the MassHunter window.

  3. Close the valve near the ESI on the MS.

  4. Set Calibrant Bottle to B at the bottom left of the Tune window.

  5. Set LC Flow to to Waste at the bottom left of the Tune window.

  6. In the Tune & Calibration tab (pictured below) set the instrument parameters required for the run, e.g.:

    • Positive – if using +ESI
    • TOF - <when would you use both?>
    • Mass Calibration / Check
    • 50-1500 m/z
    View of the Tune & Calibration

    The Tune & Calibration tab

    From Agilent’s manual:

    Before you calibrate the instrument, you have to set the instrument state
    to the proper instrument mode, mass range and fast polarity switching
    mode. You set these values on the Instrument State tab.
    
    When you change the mass range or enable/disable fast polarity switching
    on the Instrument State tab, the pulser frequency is changed which results
    in the DEI pulser warming up or cooling down. If the calibration is
    performed too soon, the DEI may still be heating up or cooling down which
    can result in drift. See the online Help for more information on the
    Instrument State tab.
    
  7. Click the button to start the calibration.

    Note

    The text of this button veries depending on the mode selected.

    • For Quadrupole, the button reads Start Check Quadrupole
    • For TOF, the button reads Start TOF Mass Calibration
    • For Both, the button reads Start Check Quadrupole & TOF Mass Calibration
  8. Once complete, the calibration report will open automatically.

  9. Under TOF Mass Calibration Data, for the largest mass check that the resolution is at least 20,000 and that the corrected residuals is below 5 ppm.

  10. On Page 2 of the report, under DetectorMCP, check that the value is below 900.

  11. Back in MassHunter, return to Acquisition mode by selecting ContextAcquisition at the top left of the window.

  12. Open the valve near the ESI on the MS.