LC-MS Checklist¶
Before the Analysis¶
Sample Preparation¶
Filter samples before injection with a
Syringe
andFilter
.- Syringe filters can be found in the cupboard in the weighing room.
- Clean syringes can be found in sealed packets in the cupboard under the
HPLC
.
Pre-Run¶
Check the indicator lights on the
Autosampler
,Pump
,Column
,VWD
andMS
are notRed
.Orange
,Flashing Orange
orOff
are OK.Check the valve near the
ESI
on theMS
is open (handle in line with pipe) and is connected toL
on the nearby block (shown below)Check that the pipe is connected to the port marked “L” to the left of this image, not to the port marked “H”
Check the levels of the bottles on the front of the
MS
. Tap the bottles to disturb the liquid if the level is not obvious.Check the levels of the waste bottles in the following locations:
- Beneath the bench below the
Pump
/Autosampler
/Eluent
stack - Beneath the bench below the
MS
and behind theVacuum Pumps
- Behind the
MS
- Beneath the bench below the
Check there is pressure in the
Collision Gas Bottle
behind theMS
, and record the pressureCheck the 3
Exhaust Pipes
are entering theExtractor
for theGraphite Furnace
Return to the front of the instrument
Fetch some blue towel.
Open the
Ionisation Chamber
of theMS
and spray50% IPA
around (but not into) the metal surfaces of the inlet orifice.Spray the inside of the
Ionisation Chamber
door and wipe with the blue towel.Close the
Ionisation Chamber
.Open Mass Hunter Workstation Data Acquisition.
Check the pressures on the top right of the screen are within the following values:
Rough Vac < 4.4×10 0 Quad Vac 1×10 −5 - 4×10−5 TOC Vac 2×10 −7 - 4×10 −7 Check sufficient eluent is present in the eluent bottles being used for the run.
Check that the
Quat. Pump
is aware of the level of eluent in the bottles. To do this:Right click on the
Quat. Pump
panel of the Instrument Status window in MassHunter. <could use image here>Select Bottle Fillings from the menu.
Under Actual Volume, ensure the value reflects the volume of eluent actually present in the bottle.
The “Bottle Fillings” window
Click OK.
Repeat the above steps when Changing Eluent (see below)
Changing Eluent¶
To change an eluent, perform the following steps:
Right click on the
QTOF
panel of the Instrument Status window in MassHunter and select LC → Waste from the menu.The Instrument Status window
In the Method Editor, under
Quat. Pump
, set the Flow Rate to 0.000 mL/min and press Enter⏎Swap the the eluent bottles. Try to swap similar eluents; for example, replace Ammonium Formate with Ammonium Acetate.
- The Lids on the bottles do not need to be tight.
Open the valve behind the door on the front of the
Pump
.Set the level of any solvents that weren’t changed to 0% and enable the eluents that were changed.
Set the Flow Rate to 1.000 mL/min and press Enter⏎
Check the pressure in the
Column
does not rise above 0 Bar. If it does, check the valve at the front of thePump
is open.Set the Flow Rate to 5.000 mL/min and press Enter⏎
After 5 minutes, set the Flow Rate to 0.000 mL/min and press Enter⏎
Close the valve on the front of the
Pump
and close the door.
A method can be setup for flushing and equilibrating the column as part of the Worklist if desired
Mass Calibration / Check¶
Select Context → Tune at the top left of the MassHunter window.
Turn on the
QTOF
by clicking the On button in its section of the MassHunter window.Close the valve near the
ESI
on theMS
.Set Calibrant Bottle to B at the bottom left of the Tune window.
Set LC Flow to to Waste at the bottom left of the Tune window.
In the Tune & Calibration tab (pictured below) set the instrument parameters required for the run, e.g.:
- Positive – if using +ESI
- TOF - <when would you use both?>
- Mass Calibration / Check
- 50-1500 m/z
The Tune & Calibration tab
From Agilent’s manual:
Before you calibrate the instrument, you have to set the instrument state to the proper instrument mode, mass range and fast polarity switching mode. You set these values on the Instrument State tab. When you change the mass range or enable/disable fast polarity switching on the Instrument State tab, the pulser frequency is changed which results in the DEI pulser warming up or cooling down. If the calibration is performed too soon, the DEI may still be heating up or cooling down which can result in drift. See the online Help for more information on the Instrument State tab.
Click the button to start the calibration.
Note
The text of this button veries depending on the mode selected.
- For Quadrupole, the button reads Start Check Quadrupole
- For TOF, the button reads Start TOF Mass Calibration
- For Both, the button reads Start Check Quadrupole & TOF Mass Calibration
Once complete, the calibration report will open automatically.
Under TOF Mass Calibration Data, for the largest mass check that the resolution is at least 20,000 and that the corrected residuals is below 5 ppm.
On Page 2 of the report, under Detector → MCP, check that the value is below 900.
Back in MassHunter, return to Acquisition mode by selecting Context → Acquisition at the top left of the window.
Open the valve near the
ESI
on theMS
.
Performing Analysis¶
Starting a Run¶
If Mass Calibration / Check was performed the LC
flow will have been set to go to the MS
automatically. Otherwise, perform the following steps to set the flow:
Right click on the
QTOF
panel of the Instrument Status window in MassHunterSelect LC → Waste from the menu.
![]()
The Instrument Status window
To start the run, in the Worklist, select the samples to run by ticking the box to the left of the Sample Name field and click the ▶ button.

The Worklist
During the Run¶
- Check that the
Column
pressure remains stable and does not rise above 200 bar. Analysis should stop automatically if the pressure rises above 600 bar. - Check that the three symbols in the
Q-TOF
panel of the Instrument Status window (shown below) remain green to indicate that the reference ions were found.

The Instrument Status window
Shutdown¶
- In the
Instrument Status
window, click the Off button to the lower right. - Close
Data Acquisition
.